Service mode

In Service mode, meter-specific operating data 
can be called up.

Activating Service mode
  1 

Hold the selection key   pressed down.

A pixel will appear in each corner of the display.

  2 

Observe one pixel:

   While the pixel is visible, hold the selection key 

 pressed down and release the key as soon 

as the pixel has disappeared.

   

Repeat the process, until all the pixels are off and 
“Test instructions” appears in the menu area.

Service mode is activated.

Service mode menu overview

Test instructions

 

Temperature

LCD pixel test

 

Test pixel screen

Cyclic test

UMI peripherals

 

UMI-1

 

UMI-2

 

UMI-3

Calibration factors

Main menu

Test instructions

Test instructions

Automatic return to main 

menu after 5minutes of

inactivity

Hold       on any screen to

return immediatelv

Temperature

The current gas temperature is displayed.

Temperature

24.39°C

LCD pixel test

A display test can be carried out in this menu.

  1 

Follow the displayed instructions.

A test pattern is shown in the display.

  2 

Briefly press the selection key  .

A further test pattern appears in the display.

   

Hold the selection key pressed down. The display 
switches to the previous menu.

Cyclic test

The accuracy of the meter can be checked using 

a cyclic volume test.

Cyclic test:  Starting

Cycles:  0-0  0000.00

C: 00.000000

m

3

U: 00.000000

m

3

Press      to abandon test

–   C = volume,
–   U = volume (temperature converted),
–   Cycles = number of complete measuring 

cycles (measuring unit rotations) – number of 
intermediate sampling points in the measuring 
cycle.

–   Overall measuring time [s].

Measurement must be carried out with a constant 
flow rate in order to ensure that the cyclic test 
produces an accurate result.

  1 

Briefly press the selection key   in order to start 
the measurement. A beep confirms the start of 
the measurement.

In order to pause the measurement, briefly press 
the selection key   again. Measurement stops 
when the full number of measuring unit rotations 
has been completed.

Measurement ends automatically after 90 min.

A beep indicates that measurement has ended.

  2 

Read off the measurement results.

UMI peripherals

The status

 

of the available interfaces for optional 

communication modules is displayed.

Summary of Contents for BK Series

Page 1: ... Only use OEM spare parts Transport On receipt of the product check that the delivery is complete see Part designations Report any trans port damage immediately Storage Store the product in a dry place Ambient tempera ture see Technical data Translation from the German D GB GB 1 Operating instructions for operators and installers Safety Contents Operating instructions for operators and installers ...

Page 2: ...5 Service cover 6 Service seal 7 Battery Type label Index plate BK G E DE 07 MI002 PTB001 002545632734 class 1 5 Q max Q min M10 0102 12345678 Installation Installing the gas meter For installing the gas meter in the pipe work refer to the operating instructions for diaphragm gas meters BK G1 6 to BK G25 and BK G1 6T to BK G25T see http docuthek kromschroeder com doclib main php language 1 folderi...

Page 3: ...n some menus may be missing The Current meter reading basic display appears when switching on the index If you are in a different menu the display will automatically change back to the basic display when no user key has been pressed for 30 s and switches off after a further 30 s You can navigate from the basic display to the various menus such as Meter information using the user keys Menu overview...

Page 4: ...nce between the internal time recording and the actual time is too large This can lead to invalid consumption data After the next time syn chronization the consumption data are recorded again correctly and disappears Tariff totals If the optional tariff data are configured they can be called up here Tariff data with measured gas volumes are dis played by pressing the selection key Previous tariff ...

Page 5: ...n the display 2 Briefly press the selection key A further test pattern appears in the display 3 Hold the selection key pressed down The display switches to the previous menu Cyclic test The accuracy of the meter can be checked using a cyclic volume test Cyclic test Starting Cycles 0 0 0000 00 C 00 000000m3 U 00 000000m3 Press to abandon test C volume U volume temperature converted Cycles number of...

Page 6: ...replaced when the optical communication link has been established see Establishing an optical communication link p 6 Otherwise data will be lost 2 Initiate changing the battery The procedure depends on your user software Attention The battery may only be removed once a release message has been issued by the index As soon as the release message NOW READY TO CHANGE THE BATTERY PACK appears there is ...

Page 7: ...nufacturer 6 Re seal the service seal Check test The MID 2004 22 EC prescribes that the measur ing instrument in this case the BK should be able to be checked The accuracy of the meter must be checked by comparing the flow rates The requirements and test methods must comply with national laws and regulations The check should be carried out using the cyclic test see Service mode p 5 Assistance in t...

Page 8: ... Optical interface pursuant to EN 62056 21 Mode E Annex B 2 For further technical data on diaphragm gas meters BK see Diaphragm meters BK G1 6 to BK G25 and BK G1 6T to BK G25T operating instructions http docuthek kromschroeder com doclib main php language 1 folderid 400041 by_class 2 by_ lang 1 Declaration of conformity Diaphragm gas meters BK E with electronic index We the manufacturer hereby de...

Reviews: