ELMASLAR LIFETIME SA-PB 1 Manual Download Page 5

1.1 Signs on device plate 

  

Symbol 

Description 

 

Serial Number 

 

 

 

Date of Manufacturing 

 

 

 

DC current 

 

 

1984 

CE mark (certifies that the device meets the requirements of 
European Union directives) 

 

Attention, please see the instructions for use. 

 

 

 

Protection class against electric shock: BF type device 

 

 

 

IP X1 Protection against water penetration 

 

 

 

Disposing this device with domestic wastes is prohibited. 

 

 

 

Store in temperatures between -40 °C and +70 °C 

 

 

 

 

1.2 Special signs on device 

  

The inscriptions on the reusable collecting container lid show the direction of the hose connections. Use the tip with the text “VACUUM” on 

the connection between the collecting vessel and the device, and the tip with the text “PATIENT” on the hose connection to the patient. In case 

of reverse connection, since the anti-overfill device will not work, secretions may enter the suction pump and damage it. To prevent the risk of 

infection, the bacteria filter between the device and the sump connection must be renewed or sterilized after each use. 

 

Capacity Indicator 

 

Warning symbol on the capacity indicator warns against the risk of over-discharging the battery pack can be damaged. You must charge 

the LIFETIME SA-PB before the red warning LED lights up or when this LED lights up at the latest. In case of not charging and continuing to 
use, the device will stop automatically when the battery level becomes insufficient and will give audible and visual warning for 6 seconds. 

 

2. Product Descriptions 

 
 

2.1 Intended Use 

  

LIFETIME SA-PB is a mobile, portable, electrically operated medical suction pump used temporarily in adults, children and infants: 

 

To aspirate blood, secretion and food particles from the oral cavity, nasal-pharyngeal region and bronchial system 

 

If LIFETIME SA-PB is used in accordance with the rules to evacuate vacuum stretchers and vacuum splints, the danger of 

respiratory arrest can be prevented by removing obstruction in the airways. This device can be used inside buildings, 

outdoors and during transport. 

 

LIFETIME SA-PB vacuum aspirator will not be used in the following areas: 

In medical condition requiring a potential balancing (for instance: cardiac surgery) 

 

In places where explosion danger is available 

 

For drainages in low vacuum intervals (for instance: wound drainages or thoracic drainages) 

 

 

2.2 Functional Descriptions 

  

An electrical pump with piston gives the vacuum pressure required for aspiration. The device can be optionally functioned with mounted 

rechargeable batteries or connected to an external direct current source of 12.0 - 13.8 V. With a vacuum setting unit, it is possible to select a 
vacuum power between –100 mmHg and -650 mmHg. Keypad is illumminated to be used in dark environment. 

 

Information 

 

When the vacuum pressure you have chosen is reached, the device switches to a constant speed. When the vacuum 

pressure changes, the pump adjusts the speed again to create the vacuum pressure selected. 

 

LIFETIME SA-PB reduces energy consumption by reducing performance when it reaches the required vacuum power. 

 
 

Page 

5

 / 

14 

Summary of Contents for LIFETIME SA-PB 1

Page 1: ...LIFETIME SA PB Suction Device Product Descriptions and Operating Manual ELMASLAR A MODELS LIFETIME SA PB 1 LIFETIME SA PB 2 LIFETIME SA PB 3 LIFETIME SA PB 4 Page 1 14...

Page 2: ...n 8 5 3 Ending Aspiration 9 5 4 Charging LIFETIME SA PB 9 6 Hygienic Preparation 9 6 1 Preparation 9 6 2 Cleansing Disinfection and Sterilization 9 6 3 Connecting Pieces 10 7 Controlling Functions 10...

Page 3: ...1 Overview Front view of LIFETIME SA PB device with reusable collecting container Back view of LIFETIME SA PB device Page 3 14...

Page 4: ...Reusable Collection Container Accessories and spare parts Page 4 14...

Page 5: ...visual warning for 6 seconds 2 Product Descriptions 2 1 Intended Use LIFETIME SA PB is a mobile portable electrically operated medical suction pump used temporarily in adults children and infants To...

Page 6: ...ing with the multi use collection container the device must always stand upright because the ball of the overfill assembly cannot safely close the connection to the suction pump and secretions may ent...

Page 7: ...template on the wall and align it with a water gage 3 Mark the holes for the assembly points 4 Make the holes by using a 3 mm sized drill bit 5 Screw the wall stabilizer by using the screws in the pac...

Page 8: ...ve integrated into suction bag by the saliva the device should always be standing vertically If the fluid is scattered onto the overflow valve it becomes air resistant and the disposable aspiration ba...

Page 9: ...cess ends all the lights on the indicator band will light up steadily Charging with wall stabilizer 1 Connect LIFETIME SA PB device to the wall stabilizer 2 Charging starts automatically a few seconds...

Page 10: ...d is 20 minutes 6 3 Connecting Pieces Following the cleaning disinfection or sterilization procedures connect the pieces as follows 1 Screw the float of the overfill prevention device into the secreti...

Page 11: ...are Correct the incorrect polar connections and reverse change the fuses if required Connection between circle plate and battery Do the interconnection entirely group is not well inserted Device is no...

Page 12: ...ck is full but the charge status of the battery pack will not be displayed unless calibration is performed 12 Do a functional check see the chapter 7 2 Function Controlling 8 1 2 Calibrating capacity...

Page 13: ...vacuum pressure in 12 V 650mmHg in sea level Aspiration hose 7 12 mm Length 1500 mm reusable container Bacteria filter Filter component of 1 m Motor power 55 W Rated voltage 12 V Maximum power 8 A Whi...

Page 14: ...ransport damage caused by improper packaging when the device is returned Not performing the required or stipulated maintenance procedures Normal erosion due to operating The following parts are under...

Reviews: