Elma Schmidbauer GmbH
16 Maintenance
Elmasonic Select
31 / 32
16.2
Clean enclosure
ü
The enclosure is dirty.
ü
The network cable is unplugged.
◦ Wipe all surfaces using a damp cloth.
WARNING! Do not spray the outside of the unit with water.
è
The enclosure has been cleaned.
16.3
Cleaning the tank
ü
The tank is full of limescale or dirty.
ü
The tank is empty.
1. Use a damp cloth to wipe out limescale and dirt residue. If necessary, rinse the tank using a
shower set.
CAUTION! Do not spray the outside of the unit with water.
ð
The tank has been cleaned.
2. To descale the tank, fill it with tap water up to the fill level marking in the tank.
3. Heat the unit up to 40 °C by pressing the
Plus
button.
4. Add 4–10% of Elma clean 60 or Elma clean 115C.
5. Once 40 °C has been reached, switch off the unit.
ð
Leave the solution to work for approx. 12 hours.
6. Then, switch on eco ultrasound mode for approx. 15 min.
7. Empty the tank.
8. Use a damp cloth to wipe out the remaining limescale and dirt residue. If necessary, rinse the
tank using a shower set.
CAUTION! Do not spray the outside of the unit with water.
è
The tank has been descaled and cleaned.
16.4
Disinfect
We recommend the following disinfectants for wiping the unit:
• Incidin Liquid by Ecolab (ready-to-use, quick disinfectant).
• Terralin protect by Schülke (observe dosage and exposure time).
◦ Disinfect the tank and the enclosure using a commercially available surface disinfectant on a
regular basis.
CAUTION! Test the disinfectant on a small area first to ensure it does not
affect the materials, particularly the controls.
è
The unit has been hygienically disinfected.
16.5
Clean the blower, and check that it is working correctly.
Dust and dirt adhering to the protective grille of the blower can impair the cooling performance or
cause unit malfunctions.
• Regularly remove accumulated dust and dirt.
ü
The unit is ready for operation and filled with fluid to enable the functions to be checked.
1. Heat the unit up to approx. 55 °C by pressing the
Plus
button.