12 Operating the unit
Elma Schmidbauer GmbH
22 / 32
Elmasonic Select
Two modes can be selected for starting ultrasound operation:
Display in Normal mode
Display in Programme mode
1
2
1
2
3
• Set the duration, ultrasound mode and
temperature.
• Switch the heating on/off.
• Set auto-degas start.
• Set temperature-controlled start.
• Save programmes.
Select or start a saved programme.
Press the Select
button
(
1
) to switch between programme mode and normal mode.
The display (
2
) shows the current settings.
The display (
3
) shows the set programme number (programmes 1–4 in Programme mode).
12.2
Set the cleaning duration
From 1 minute to 6 hours.
Press one of the
Plus/Minus
buttons below the time display to set the duration. Press the
button repeatedly until the required duration is shown on the display. Hold the button
down to increase/decrease the duration more quickly.
12.3
Set the target temperature
From 25–80 °C/77–176 °F.
Press one of the
Plus/minus
buttons below the temperature display to set the required
target temperature. Press the button repeatedly until the required temperature flashes on
the display. Hold the button down to increase/decrease the duration more quickly.
IMPORTANT
If a temperature limit between 40–60 °C/104–140 °F has been set, see the
Change parameters section for information on changing the temperature limit.
Longer periods of ultrasound operation heat the cleaning fluid in addition to the heating
function. This can lead to the fluid being hotter than the set target temperature.
12.4
Switch the heating on/off
Press the
Plus
button below the temperature display to switch the heating on. The
Heating
symbol appears on the display when the heating is switched on. The display
shows the set target temperature and the current actual temperature of the cleaning fluid
flashing alternately.