BM869s
50
_v61010-2-033:2019
Electrical Specifications
Accuracy is
(% reading number of digits) or otherwise specified, at 23
C
5
C &
less than 75% relative humidity.
True RMS voltage & current accuracies are specified from 5 % to 100 % of range or
otherwise specified. Maximum Crest Factor < 2.25:1 at full scale & < 4.5:1 at half scale,
and with frequency components within the specified frequency bandwidth for non-
sinusoidal waveforms.
DC Voltage
RANGE
BM869s
BM867s
Accuracy
500.00mV,
5.0000V,
0.02% + 2d
0.03% + 2d
50.000V
0.03% + 2d
0.04% + 2d
500.00V
0.04% + 2d
0.05% + 2d
1000.0V
0.15% + 2d
0.15% + 2d
Input Impedance: 10M
, 60pF nominal (80pF nominal for 500mV range)
Ohms
RANGE
BM869s
BM867s
Accuracy
500.00
0.07%+10d
0.1%+10d
5.0000k
0.07%+2d
0.1%+6d
50.000k
0.1%+2d
0.1%+6d
500.00k
0.1%+2d
0.1%+6d
5.0000M
0.3%+6d
0.4%+6d
50.000M
2.0%+6d
2.0%+6d
99.99nS*
2.0%+10d
2.0%+10d
Open Circuit Voltage: < 1.3VDC ( < 3VDC for 500
range)
*From 0% to 10% of range: Specified ac 30d
Audible Continuity Tester
Audible threshold: between 20
and 200
Response time < 100
s
Crest mode (Instantaneous Peak Hold)
Resolution: 5000 counts
Accuracy: Specified accuracy
100 digits for changes > 0.8ms in duration