AM852 Veterinary Anaesthesia machine User’ Manual
23
6. Cleaning and Sterilization
6.1Cleaning and sterilization of Anaesthesia workstation surface
Clean the machine's panel and all surfaces with soft cloth soaked in the common water
soluble disinfectant. The confection of the disinfectant must be done in accordance with
the directions given by the manufacturer. One must prevent the disinfectant drops from
entering the Anaesthesia apparatus and the misusage of organic solvent for cleaning the
machine.
6.2Cleaning and sterilization of reusable rubber corrugated tube and
manual breathing bag
Every used rubber corrugated tube and manual breathing bag must be firstly washed with
rinsing carefully and after they were dried put them into the suffocating sterilization
chamber to be suffocated and sterilized for a while or sterilize them as the rubber products
require to do. Don’t sterilize them with ultraviolet radiation, or else it will result in the fast
aging of rubber products.
Note:
Do not reuse one-off consumable.
Warning:
Ensure the avirulence of corrugated hose and respiratory bag, and the followings
must not emerge:
·
Cause allergic reactions in animal’s body.
·
React with Anaesthesia gas or anesthetic to produce dangerous byproducts.