background image

CLD-100
HALOGEN LEAK DETECTOR

Instruction Manual

General Information

 Incorporated  with advanced leak detection technology, CLD-100 is

Detect all kinds of halogen refrigerants

 an ideal and economic halogen leak detector.Compact in appearance, 

 it is stable and easy to operate.

Sensitivity can be adjusted at any time and the detector can be adjusted    

automatically to the best state of detection.

Inner precise IC with ultra-low circuit consumption offers more stable 

function and lengthens battery life.

Two-color visual display of battery voltage.

Excellent sensor brings high sensitivity and longer service time.

General Information

Detection Range

 

Battery lndicator

The forepart light of panel indicates the leak and the battery voltage.

GREEN       Battery voltage is normal, sufficient for proper operation.

ORANGE    Battery voltage is approaching the lower threshold for 

operation.

Replace them as soon as possible.

 

CFCs       e.g.R12,R11,R500,R503 etc.

 

HCFCs    e.g.R22,R123,R124,R502 etc.

  HFCs       e.g.R134a,R404a,R125 etc.

  Blends such as AZ-50, HP62, MP39 etc.

Detect Ethylene Oxide gas leaks in hospital sterilizing equipment

(it will detect the halogenated carrier gas).

Detect SF-6 in high voltage circuit breakers.

Detect most gases that contain Chlorine, Fluorine and Bromine(ha

logen gases).

Detect cleaning agents used in dry cleaning applications such as 

perchloroethylene.

Detect halogen gases in fire-extinguishing system.

 

Operating temperature: 

 

Maximum sensitivity:

  Battery life: 

  Response time:

  Work mode:

  Fixed probe length: 

  Warm-up time: 

  Reset time: 

  Power supply:

0

 to 52

(30

 to 125

)

6 gr/yr, for all halogenated refrigerants

approximately 20 hours in normal use.

instantaneous.

continuous, no limitation.

20cm.

approximately 6 seconds.

2-10 seconds.

6V DC, four AAA excellent batteries.

The CLD-100 may also be used to detect leaks in other systems and storage

/recovery containers.It will respond to all halogenated (including Chlorine and 

Fluorine) refrigerants. This includes but is not limited to: