Appendix
2: ELCORplus version for Zone 2
194
5 Product features
5.1 Metrological characteristics and parameters
Metrological characteristics and parameters of device ELCOR
plus
for Zone 2 are identical to the
standard version of the intrinsically safe ELCOR
plus
device.
5.2 Technical design and characteristics
Technical design and characteristics of the device ELCOR
plus
for zone 2 are almost no different
from intrinsically safe but a modified version of the I/O board and the associated changes in plug-
in modules are used.
Some terminals for connecting external devices are not intrinsically safe on this unit and have a
“nA” type of protection. This changes the way of connecting some additional devices
(communication devices, digital outputs and possibly analog inputs / outputs).
These changes are described in the following chapters.
Summary of Contents for DATCOM
Page 38: ...3 Technical Parameters 36 ...
Page 50: ...5 Installation and Commissioning 48 Fig 7 Examples of data plates ...
Page 83: ...5 Installation and Commissioning 81 Fig 37 Communication via RS232 RS485 using B RS module ...
Page 115: ...7 Technical Description of the Device 113 Fig 57 Functional diagram of ELCOR ...
Page 116: ...7 Technical Description of the Device 114 Fig 58 Functional diagram of ELCORplus ...
Page 122: ...7 Technical Description of the Device 120 Fig 61 Modem replacement ...
Page 159: ...10 Communication with the Device 157 packet ...
Page 174: ...12 Accessories 172 Fig 78 ELCORplus example of use of external modules ...
Page 213: ...Appendix 3 Compass Data Export 211 Preview of generated document ...
Page 215: ...Appendix 3 Compass Data Export 213 3 Export to XLS 4 Export to TXT DOC ...
Page 216: ...Appendix 3 Compass Data Export 214 ...
Page 217: ...Appendix 3 Compass Data Export 215 5 Export of graph ...