Visual field tests
© 2017 Elektron Technology
3 .2 .3 .1 Me a n De fe ct
The Mean Defect score (w hich is given on the printout) gives the clinician an
estimate of the extent of damage.
A more negative value indicates a less sensitive eye. Mean Defect values are
sensitive to both scotomata and media opacities. The value w ill go dow n if a
patient has a glaucomatous field defect and if they have a cataract.
As the name implies, the Mean Defect score is based on the average of all the
values, excluding those from the blind spot area. How ever, as some
test locations are more variable than others, it is w eighted to give more
importance to the less variable locations (those nearer the centre of the tested
field).
A P-value is given w hen the probability of the MD value coming from a normal
patient is below 10% (possible outputs are <10%; <5%; <2%; <1%; <0.5%).
When it is above 10% then the message
W ith in N o rma l L imits w ill be displayed.
T he f ig ur e a b o v e s ho ws the g lo b a l ind ic e s a s a p p e a r o n the s ta nd a r d p r int
o ut.
51
Summary of Contents for HENSON 8000
Page 1: ... 2017 Elektron Technology Henson Perimeter User Manual ...
Page 6: ......
Page 7: ...Part I ...
Page 14: ......
Page 15: ...Part II ...
Page 21: ...Part III ...
Page 25: ...25 Visual field tests 2017 Elektron Technology ...
Page 41: ...41 Visual field tests 2017 Elektron Technology ...
Page 64: ......
Page 65: ...Part IV ...
Page 85: ...Part V ...
Page 95: ...Part VI ...
Page 106: ......
Page 107: ...Part VII ...
Page 116: ......
Page 117: ...Part VIII ...
Page 121: ...Part IX ...
Page 123: ...Part X ...
Page 127: ...Part XI ...
Page 136: ......
Page 137: ...Part XII ...
Page 146: ......
Page 147: ...Part XIII ...
Page 149: ...Part XIV ...
Page 156: ......