Visual field tests
© 2017 Elektron Technology
3 .1 .7
C h a n g in g t h e s u p ra -t h re s h o ld in cre me n t
There are 3 different Supra-threshold test increments, 5, 8 and 12dB.
Each of the test increments has a button on the icon bar (see above). The
button that appears to be dow n gives the currently selected increment. The
increment can be changed by clicking over the required button .
Single stimulus
In the Single stimulus supra-threshold test the increment is adjusted
automatically. If a stimulus is missed the program w ill come back to the location
and re-test it at the same intensity. If it is missed a second time it w ill come
back to test it at 8dB above threshold and if missed at 8dB then at 12dB above
threshold.
Multiple stimulus
In the multiple stimulus supra-threshold test the program starts off testing at
5dB above the threshold estimate. If the patient reports less than the correct
number of stimuli then the pattern should be repeated. If they still report less
than the correct number then the patient is asked to report w here they saw
the stimuli. The perimetrist then
(s) they missed. The
perimetrist should then alter the supra-threshold increment in order to
measure the depth of the defect.
36
Summary of Contents for HENSON 8000
Page 1: ... 2017 Elektron Technology Henson Perimeter User Manual ...
Page 6: ......
Page 7: ...Part I ...
Page 14: ......
Page 15: ...Part II ...
Page 21: ...Part III ...
Page 25: ...25 Visual field tests 2017 Elektron Technology ...
Page 41: ...41 Visual field tests 2017 Elektron Technology ...
Page 64: ......
Page 65: ...Part IV ...
Page 85: ...Part V ...
Page 95: ...Part VI ...
Page 106: ......
Page 107: ...Part VII ...
Page 116: ......
Page 117: ...Part VIII ...
Page 121: ...Part IX ...
Page 123: ...Part X ...
Page 127: ...Part XI ...
Page 136: ......
Page 137: ...Part XII ...
Page 146: ......
Page 147: ...Part XIII ...
Page 149: ...Part XIV ...
Page 156: ......