Elektro-Automatik 39 200 120 Operating Manual Download Page 26

Page 26

EA Elektro-Automatik GmbH

Helmholtzstr. 31-37 • 41747 Viersen

Germany

Fon: +49 2162 / 3785-0

Fax: +49 2162 / 16230

www.elektroautomatik.de

[email protected]

PS 2000 B Triple Series

3.5 

Remote control

3.5.1  EA Power Control

The included USB stick contains the Windows software EA Power Control which can remotely control the device. 

Also see the user manual of the software.

3.5.2  Programming

Programming details and the communication protocol are to be found in the programming documentation which 

is supplied on the included USB stick or which is available as download from the EA Elektro-Automatik website.

3.5.3  EasyPS2000

The Windows software EasyPS2000, as offered for the former PS 2000 B design generation with blue LCD, is 

still available and can also be used with the new design generation with color TFT, but it would still depict the 

counterfeit of the old design. 

3.6 

Alarms and monitoring

3.6.1  Device alarm and event handling

A device alarm incident will usually lead to DC output switch-off and the appearance of a 

text message in the display to make the user aware. Some alarms must be acknowledged.  

If the condition persists, the display remains and the alarm can only be acknowledged 

after elimination of the cause. 

 

How to acknowledge an alarm in the display (during manual control)

1. 

Once an alarm is indicated, the user can try to acknowledge and delete the alarm by pressing button 

On/Off

.

In order to acknowledge an alarm during remote control, see the programming guide.

These device alarms are configurable:

Alarm Meaning

Description

Range

Indication

OVP

O

ver

V

oltage 

P

rotection

Triggers an alarm if the DC output voltage reaches the 

defined threshold. The DC output will be switched off..

0 V...1.1*U

Nom

Display, digital 

interface

OCP

O

ver

C

urrent 

P

rotection

Triggers an alarm if the DC output current reaches the 

defined threshold. The DC output will be switched off..

0 A...1.1*I

Nom

These device alarms can’t be configured and are based on hardware:

Alarm Meaning

Description

Indication

OT

O

ver

T

em-

perature

Triggers an alarm if the internal temperature exceeds a certain limit. 

The DC output will be switched off.

Display, digital 

interface

 

► How to configure the device alarms

1. 

While the DC output is switched off, press button 

Settings

.

2. 

In the menu use the left knob to navigate to 

Settings

 and press the knob.

3. 

In menu 

Settings

 use the left knob again to select 

OVP

 (threshold of the overvoltage protection) or 

OCP

 

(threshold of the overcurrent protection). Adjust the value as desired.

These protection threshold are always adjustable in their full range of 0...110% of the rated val-

ue. The device permanently compares the actual voltage and current on the DC output against 

these thresholds and thus an OVP alarm could also occur while the DC output is switched off. 

In this situation the voltage that triggers the OVP must come from outside and could damage 

the device.

Summary of Contents for 39 200 120

Page 1: ...Elektro Automatik PS 2000 B TFT Triple DC Power Supply Operating Manual Doc ID PS2TEN Revision 14 Date 11 2020...

Page 2: ......

Page 3: ...res before installation and use 14 2 3 2 Preparation 14 2 3 3 Installing the device 14 2 3 4 Connection to DC loads 15 2 3 5 Grounding of the DC outputs 15 2 3 6 Connecting the USB port 15 3 OPERATION...

Page 4: ...as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for general safety notices instructions and damage protection bans S...

Page 5: ...B Construction Version B Second generation Maximum current of the device in Ampere Maximum voltage of the device in Volt Special marking 3 Three outputs triple Series 2 Series 2000 Type identification...

Page 6: ...ulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device is switched off It s not necessary...

Page 7: ...he responsibilities for operation maintenance and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularl...

Page 8: ...pair of DC output sockets and is called output 2 The upper pair of DC output sockets form output 3 The meaning of the signals on the display of output 1 and 2 is as follows Signal OT OverTemperature...

Page 9: ...P 24 mVRMS Output 1 2 Current Adjustment range 0 6 A 0 10 A 0 3 A 0 5 A Accuracy 1 at 23 5 C 0 3 IMax 0 3 IMax 0 3 IMax 0 3 IMax Load regulation at 0 100 UOUT 0 15 IMax 0 15 IMax 0 15 IMax 0 15 IMax O...

Page 10: ...Automatik GmbH Helmholtzstr 31 37 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 2000 B Triple Series 1 8 4 Views Figure 1 Front side...

Page 11: ...alue or adjustment of parameter value in the menu Push selection of the decimal position cursor of the currently assigned value 7 Pushbutton On Off Used to switch the DC output oon or off unless the H...

Page 12: ...es For these devices the following accessories are available Multi Control license Ordering numbers Single license 33 100 229 5 pack of licenses 33 100 230 The device is delivered with an USB stick th...

Page 13: ...tatus display lower part This area displays various status texts and symbols Display Description The HMI is locked The HMI is unlocked Remote The device is under remote control via USB Alarm xxx Alarm...

Page 14: ...3 1 Safety procedures before installation and use Before connecting to the mains ensure that the connection is as shown on the product label Overvoltage on the AC supply can cause equipment damage In...

Page 15: ...ding of the DC outputs The metal socket in proximity to output 1 is internally wired to PE and can be used to ground one of the DC output poles of all three outputs by direct connection or it can be u...

Page 16: ...e Voltage regulation is also called constant voltage operation CV The DC output voltage of a power supply is held constant on the adjusted value unless the output current reaches the adjusted current...

Page 17: ...will be in constant voltage mode If however the actual current reaches the set value of current the device will switch automatically to current limiting While the DC output is switched on and constan...

Page 18: ...ltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage higher than set for the overvoltage alarm threshold OVP 0 110 UNo...

Page 19: ...or a blackout doesn t leave the device enough time to store the latest values It s practically powered down immediately Thus the device stores the set values and all settings automatically when they...

Page 20: ...left HMI Element Description In right HMI Language Selection of the display language between German and English Default setting English No Backlight The choice here is whether the backlight remains Al...

Page 21: ...bs 1 With the main display being shown it means no menu mode is active turn the left hand knob to adjust the output voltage and the right hand knob to adjust the output current This doesn t depend on...

Page 22: ...How to lock the HMI with PIN Don t activate the PIN lock if you are unsure about the current PIN It can be changed but only if the current PIN is entered Resetting the device via remote command will a...

Page 23: ...ivated tracking is saved along with other device settings and thus instantly restored after powering the device Even in tracking mode the two outputs can have separate alarms which would always lead t...

Page 24: ...t used the serial current is limited by the output with the lower current set value Series connection of multiple devices isn t permitted because the total voltage could exceed the voltage withstand o...

Page 25: ...considered completely differently to the other outputs It s meant to be an auxiliary output with a fixed voltage which is barely adjustable ca 3 to ca 6 V with a trimmer That trimmer is placed between...

Page 26: ...can try to acknowledge and delete the alarm by pressing button On Off In order to acknowledge an alarm during remote control see the programming guide These device alarms are configurable Alarm Meani...

Page 27: ...detail as possible is attached if shipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by a fusible which is inside a f...

Page 28: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 37 41747 Viersen Germany Fon 49 2162 37 85 0 Telefax 49 2162 16 230 ea1974 elektroautomatik de www elektroautomatik de...

Reviews: