Page 2: ...Operation Method 6 1 Power On Off 6 2 Battery Voltage Check 6 3 Precision Measurement of Equipotential Resistance 6 4 DC Voltage Test 7 5 AC Voltage Test 7 6 Backlight Control 8 7 Alarm Setting 8 8 Data Lock Storage 8 9 Data Review Deletion 8 10 Data upload 9 VII Battery Description 9 VIII Accessories 9 I Safety Precautions and Procedures ...
Page 3: ...clean do not drop u Use disassembly and maintenance of this tester must be performed by authorized personnel u Because of the reason of this tester if it is dangerous to continue using it it should be stopped immediately and sealed immediately and handled by an authorized institution u The hazard symbol on the tester and manual the user must be safely operated according to the indication II Introd...
Page 4: ...0 0mΩ 1 rdg 10dgt 0 1mΩ 301mΩ 3000mΩ 1 rdg 5dgt 1mΩ 3 01Ω 30 00Ω 0 01Ω 30 1Ω 300 0Ω 0 1Ω 301Ω 3000Ω 1Ω 3 01kΩ 30 00kΩ 10Ω DC voltage 0 1 100 0V AC 1 5 rdg 5dgt 0 1V 101 1000V 1V AC voltage 0 1 100 0V AC 1 5 rdg 3dgt 0 1V 101 750V 1V Remark 23 5 below 75 rh IV Technical Specifications Function Equipotential bonding resistance test between metal components low value resistance test resistance test o...
Page 5: ...ULL symbol indicates that the storage is full Data review Data review function MR symbol display Overflow display Over range overflow function OL symbol display Alarm function Alarm when the measured value exceeds the alarm setting value Battery voltage Real time display of battery power reminding timely charging when battery voltage is low Automatic shut down APO indicates automatic shutdown afte...
Page 6: ...op resistance tester and DC resistance fast Tester verification procedure V Structure 1 LCD 2 C1 interface positive current 3 P1 interface positive voltage 4 P2 interface negative voltage 5 C2 interface negative current 6 Function button 7 DC Charging stand 8 USB interface 9 Test button 10 Power on off button 11 Safety alligator clip 12 Test line red black each 2PCS VI Operation Method 1 Power On ...
Page 7: ...The connection between the test line and the detector and the measured object should be reliable Cause the component test line is heated it will cause errors It is recommended that the test result be more accurate every 30 seconds During the test the detector displays the OL symbol indicating that the equipotential value between the two points measured exceeds the range check the test line contact...
Page 8: ...the red test line and the V end P1 terminal of the meter to the other end of the measured object s the positive pole The black test line and the COM terminal P2 is connected to the negative pole of the measured object The LCD displays the real time DC voltage value 5 AC Voltage Test Input meter AC voltage can not exceed 750V When measuring AC voltage only need to connect P1 and P2 terminals Do not...
Page 9: ...n the measured voltage value is greater than the alarm critical set value or the insulation resistance value is less than the alarm critical set value and the alarm function is turned on the instrument flashes the symbol and issues a beep beep beep alarm sound 8 Data Lock Storage After the measurement is completed short press MEM key to lock the current display data and automatically serial storag...
Page 10: ...ve it The monitoring software has online real time monitoring and history query function dynamic display alarm value setting and alarm indication function with historical data reading reviewing saving printing and other functions VII Battery Description Generally charging 8 to 10 hours If the detector is not used for a long time please charge the battery once every 1 2 months Please use the origin...
Page 11: ... losses caused by the use The company reserves the right to modify the contents of the user manual Any changes will be made without further notice GuangZhou ZhengNeng Electronics Technology Co Address 2F No 15 Baoshu Road Taihe Baiyun District Guangzhou Guangdong China Toll free call 4000 1515 38 Tel 86 20 36544172 Fax 86 20 37319075 Post 510540 WebSite www znele com ...