Calibration Guide 

 

 

RH1 RH2 
RH5 RH6

 

 

 

 

Summary of Contents for Check-Line RH1

Page 1: ...Calibration Guide RH1 RH2 RH5 RH6 ...

Page 2: ...device with the 80 humidity standard The time for conditioning is two hours Switch on the device and annotate the shown measurement value If the difference between the shown value and the calibration value is bigger than 2 0 rh we advice you to recalibrate your RHx Navigate to the menu Calibrate Ic V assume the active IC Calibration value of the 80 and annotate this Don t push any key until the de...

Page 3: ...sket ring in the bottom part of the case Attach the metal ring on the third gasket Take the upper part with the RHx and attach these carefully at the bottom part of the calibration device Screw it up like shown in the picture Pick up the RHx togheter with the calibration device STRAIGHT and DON T TURN IT AROUND For RH1 and RH2 place the textile pad in the bottom Part pour the humidity standard car...

Page 4: ...o this as following Ascertain the IC values You have to ascertain the IC calibration values both with the 35 rh liquid and with the 80 rh liquid 1 Let your RHx in the calibration device and switch on the device 2 Press the rolling menu key as often until you get to the main menu 3 Choose the Options menu by pressing and confirm this with 4 Go to the Menu unlock and confirm with 5 Enter the SuperUs...

Page 5: ...to Calibrate by pressing the key and confirm it with 6 Next go to the menu I V and confirm with Now the 1st index Idx 1 will be shown 7 Now the 1st index Idx 1 will be shown on display To enter the new Ic value press the key until you get to the correct number and confirm this with If the number is taken the coursor jumps to the next number Like that you enter the whole Ic value for this Index At ...

Page 6: ...your SuperUser password and confirm it with 5 The SuperUser password if not changed is the serial number of the instrument which is shown at the start up logo and in the menu Status 1 Navigate to Reset by pressing the key and confirm it with On the display the query Reload appears on the display reset erscheint am Display 2 Confirm this query by pressing the key and the original calibration will b...