© 2020, Electrolab, Inc.
March 2020
Configuration Request Commands
Report Number of Floats
UuuF?
Response: UuuFn
uu = unit number
n = number of floats (1 or 2 – Standard; 11 or 12 – 1/8” Resolution; 11 – Single Float; 12 –
Dual Float).
Report Level Offsets
UuuLO?
Response: UuuL1Osnn.nnL2Osnn.nn
uu = unit number
snn.nn = sign and offset with two implied decimal places
Report Temperature Offset
UuuOF?
Response: UuuOFsff
uu = unit number s = sign
ff = temperature offset (degrees F)
Report Multiple Temperature Offsets
UuuTO?
Response: UuuTnOs0.0
TnOsn.n repeated for additional temperature sensors
uu = unit number
n = temperature sensor (1-8, 1 is top sensor)
s = sign
0.0= temperature offset
Report Switch Distance
UuuD?
Response: UuuDd
uu = unit number
d = distance between switches as integral tenths of an inch (e.g., 5 = 0.5 inches, 10 = 1
inches)
UuuR?
Response: UuuRmmm
uu = unit number
mmm = delay
Report Total Switches
UuuS?
Response: UuuSssss
uu = unit number
ssss = total number of switches in the sensor