a. Assigned Channel - choose from settings.
b. Watt-hour per Pulse - enter the value in the secondary.
c. Pulse Width - set from 5ms to 635ms.
d. Mode - select to enable or disable.
e. Form - select between Form A and Form C.
2. Click
OK
and then click
Update Device
to send the new settings to the meter.
NOTE:
The unit for the calibration pulses is watt hour. The initial value of each pulse
(Kt) is 1.800 secondary watt hour per pulse.
4.1.2: LCD Display
In Test Mode, the LCD display provides user access to accumulated real, reactive and
apparent energy as well as Block Demand for real power. There are eight screens in
Test Mode: Wh - Del, Wh - Rec, VARh - Q1 & Q2, VARh - Q3 & Q4, VAh - Q1 & Q4, VAh
- Q2 & Q3, Watt Demand Test Rolling and Watt Demand Test Block. See Section 4.2.2
for more detail.
The LCD display also includes three phase Voltage indicators. These indicators are
labeled A, B and C to represent the presence of Voltage on the labeled phase.
If Voltage is not present on one phase, its indicator is not displayed. These indicators
also show whether current is flowing on the phase. If no current is flowing on a phase,
the indicator blinks; when current is flowing, the indicator shows a continuously
illuminated digit. The indicator also shows phase sequence.
4.2: Test Mode
Test Mode allows the meter to be tested without disturbing billing data or setting a
new maximum Demand. Test Mode performs the same function as setting the point-
ers back on an electromechanical meter after testing.
While operating in Test Mode, the meter uses the same measurement and calculation
processes that are used in Normal Mode. The only difference is that the billing
numbers stored in the meter are not updated with the real-time reading. This
preserves the billing numbers while testing is performed.