Safe working load cannot be exceeded.
Electric adjustment functions can be used only by operators, who know
all rules related to safe operating of adjustable nursing beds.
Before every use make sure that the nursing bed is in proper and faultless
condition.
Use only technically perfect bed.
Contact manufacturer or distributor or service personnel if the exchange
or repair of faulty parts have to be made.
If damage or malfunction is suspected, take the bed out of service
immediately, label it as non-conforming product and disconnect it from
the power until exchange or repair of broken components is made.
Use only original Elbur accessories
–
modification of nursing beds
is strictly forbidden without authorization of our company.
It is necessary to perform specific control tests after modifying the
nursing bed to make sure that it is going to work properly.
Modification of nursing bed without authorization of Elbur company can
caused hazards for patients ad operators, for example entrapment
between side rails or inadvertent patient falls.
Make sure that other attached equipment can function in a secure
and safe manner (do not hesitate to contact manufacturer with
questions).
Make sure that all castors have their brakes applied when patient
is on the nursing bed and location of the bed is not going to change.
Before adjusting the height of mattress support platform, make sure
there are not any obstacles, such as furniture, window sill etc., which can
stop high/low motion.
Do not use area under the bed as a storage space.
The special attention should be given to children
–
keep the hand switch
away from them.
2.4.2. Electrical components information
Warnings
Connect the bed directly to an electrical outlet
–
do not use extension
cords or multiple outlet extension cords.
The nursing bed should be located in position which simplifies quick
and safe disconnect of mains plug from the socket.
It is obligated to perform periodic inspections and electrical tests
according to EN 62353 performed by technically trained person.
Do not open any electric components.
For reasons of safety and electromagnetic compatibility, use only original
electric components from Elbur that have been released for the nursing
bed model in order. Otherwise it can caused danger for patients and loss
of bed functionality.