12. INPUT/OUTPUT TEST STEPS
12.6.2
Start /execute test plan (step mode)
Note:
Test steps are running in "step" mode. All steps must be started by
the operator.
12.6.2.1
Read digital input
Press the green "Run" - button, for start the read action.
In this case, the tester reports "Error: digital inputs" - forecasting: 1 signal
at input 1, specified "0" in the parameter block..
See also chapter 10.5.3 Error handling:
The tester is waiting for user input.
12.6.2.2
Read Analog Input
Press the green "Run" - button, for start the read action.
In this case, the tester reports "Error: Min value underrun" - forecasting:
min. value of 100 V specified in the parameter block.
See also chapter 10.5.3 Error handling.
The tester is waiting for user input.
State Digital Inputs
Values Analog Inputs
State Digital Inputs
Values Analog Inputs
Fig. 12.27: Read digital input
Fig . 12.28: Read analog input
Full-electronic G1-1 HV DC-Tester