dilatometer. Creeping of the working electrode is induced each time the mechanical
properties of the working electrode are altered by charging. Therefore, each charge induced
height change is followed by some creeping. The charge induced creeping effects are real and
not artefacts of the measurement.
9 EC-Link Software Installation
In order to record the displacement signal together with the cell voltage, cell current,
electrode potential and temperature, the software of the integrated data logger needs to be
installed on a Microsoft Windows
®
PC.
You must be logged into an account with administrator privileges.
Save your work and close down all active programs.
On the installation media, run
CDM*_Setup
. This will install the FTDI driver required to
establish the USB connection with the data logger.
On the installation media, run
Setup
. This will install the EC-Link data logger software.
Follow any instructions that may appear on your screen.
Once installation is finished plug in the provided USB cable into both the host PC and the
ECD-3-nano controller box.
Launch the EC-Link data logger software if not already done.
After a few seconds, EC-Link should report a valid connection and you are ready to start
the measurement.
Additional information on the EC-Link software can be found in a separate manual
https://el-cell.com/support/manuals
).
10 Calibration and Settings
Calibration of the instrument has been carried out at the factory. The corresponding settings
of the EC-Link software are stored in the file
Settings ECD-3-nano [Device-ID].V2
in the
installation directory on the local hard drive and on the installation media. If the default
settings have been changed for any reason, the original settings can be restored by copying
Settings ECD-3-nano[Device-ID].V2
from the installation media into the directory
C:\ProgramData\EC-LINK. The settings affect only the data logger readings.
The DIP switches at the controller box affect both the displacement readings of the USB data
logger and the analog output signal. Leave the switches in their default position as shown in
the screenshot on the next page.