
Troubleshooting
AIM2505A Technical Manual V2.03
38
6.6.4 Command ser - serial link diagnostic counters
This command prints the I/O bus diagnostic counters for this module.
>ser
receivedok:
11190254
unknowncerr:
0
crcposerr:
0
crcsumerr:
0
framelenerr:
0
idletimeout:
0
If any of above values that have err at the end are non-zero, it means that the AIM
has experienced transmission errors on the I/O bus. This can be an indication of a
problem with the AIM module or with a CPU module.
6.6.5 Command freq
– print status of Frequency Inputs
This command prints the measured frequencies and accumulated pulse counter
values of
all four frequency inputs. Example:
>freq
1:
1000.00Hz 347412
2:
400.00Hz 17684
3:
0.00Hz
0
4:
0.00Hz
0
>freq
1:
1001.00Hz 349869
2:
399.00Hz 18667
3:
0.00Hz
0
4:
0.00Hz
0
Above example shows that the frequency input 1 sees a frequency of 1000 Hz and
1001 Hz in successive readings and the frequency input 2 sees 400 Hz and 399 Hz.
The pulse counter values can be seen incrementing after successive commands.
6.6.6 Command in [num] - print input channel values
When given the num parameter (num=1...10), this command prints the status of a
single input. Without num, prints status of all 10 channels.
>in
1:
-5.010V
0.010mA
2:
0.000V
0.010mA
3:
3.960V
0.010mA
4:
-0.040V
-0.020mA
5:
-0.020V
0.010mA
6:
-0.040V
-0.020mA
7:
8.050V
0.010mA
8:
0.000V
0.070mA
9:
-0.010V
0.030mA
10:
0.001V
16.050mA
In the example above (values in bold for emphasis), different voltages were supplied
to the inputs 1, 3, and 7. Input 1 sees 5.010 V, input 3 sees 3.960 V, input 7 sees
8.050 V. Current input 10 was supplied 16mA from a calibrator and reads as 16.050
mA. These printed voltage values should be close to ones measured with a
multimeter. Rest of the inputs were not connected and show some values near zero
volts and milliamperes.