18
Chapter3Troubleshooting
Problem
Possible cause and remedy
• Themessageshowsthattheinputsignalis
outofthespecifiedfrequencyrange.
Example:
fD: Dotclock
fH: Horizontalscanfrequency
fV: Verticalscanfrequency
• CheckwhetherthePCisconfiguredtomeettheresolution
andverticalscanfrequencyrequirementsofthemonitor
(see“2-1.CompatibleResolutions”(page13)).
• RebootthePC.
• Selectanappropriatesettingusingthegraphicsboard’s
utility.Refertothemanualofthegraphicsboardfor
details.
●
Error code table
Error Code
Description
0***
• ErrorsthatoccurredduringSelfCalibration
1***
• ErrorsthatoccurredduringGrayscaleCheck
*1**
• ErrorsthatoccurredduringDICOM
*2**
• ErrorsthatoccurredduringCustom
*3**
• ErrorsthatoccurredduringCAL1
*4**
• ErrorsthatoccurredduringCAL2
*5**
• ErrorsthatoccurredduringCAL3
**10
• Theproductmaximumbrightnessmaybelowerthanthetargetbrightness.
• Lowerthetargetbrightness.
**11
• Theproductminimumbrightnessmaybehigherthanthetargetbrightness.
• Raisethetargetbrightness.
**34
• Thesensormaynothavecomeoutduringcalibration,orlightmayhaveenteredthe
sensor.
• TurnofftheACadaptermainpower,waitforafewminutesbeforeturningthepower
backon,andthenexecuteSelfCalibration/GrayscaleCheckagain.
**61
• Thesensormaynothavecomeout.
• Checkwhetherthereisanyforeignobjectnearthesensor.
• ExecuteSelfCalibration/GrayscaleCheckagain.
**94
• SelfCalibration/GrayscaleCheckcannotbeexecutedduetothefollowingsettings:
-Thereisnosignalinputtotheleftscreen,andtheverticalresolutionoftheright
screenislowerthan2048dots.
-Theverticalresolutionsoftheleftandrightscreensarelowerthan2048dots.
**95
• TheexecutionconditionsforSelfCalibration/GrayscaleCheckmaynotbeappropriate.
• Checktheexecutionconditionsandresetthemifnecessary.