Chapter4ControllingMonitorQuality
4-4. Setting On/Off for Warning and QC History
This function allows you to set On/Off for Warning and QC History when the Grayscale Check result is Failed.
Procedure
1. Choose “RadiCS SelfQC” from the Adjustment menu, and choose
.
2. Choose “Settings” from the “RadiCS SelfQC” menu, and choose
.
3. Select “Warning” or “QC History” with
or
.
4. Select “On” or “Off” with
or
.
5. Choose
to exit.