12
WWW.EILERSEN.COM
The status automatically returns to
LOCKED
after 5 minutes without keyboard activity in
NORMAL
screen or when the terminal is power-cycled.
–
Enter values
Perform these steps to enter a new parameter value:
•
Unlock parameters as described in the above section.
•
Use the menus to select the screen where the value is displayed.
•
Use and/or to highlight the parameter.
•
Press
to start entering a new value.
Entry using selection list
Some parameters (such as resolution and decimal point position) are entered using a selec-
tion list. When change of this type of parameter is requested, a special pull-down menu will
appear with a list of possible values as shown:
The keys can be used as follows:
F
Move the cursor down in this selection list
.
Move the cursor up in this selection list.
Move the cursor down in this selection list.
Accept the selected/marked value as the new desired value.
Clr
Abort the entry without change of parameter. This can also be done by selecting
the “CANCEL” entry form the selec
tion list.
Example - Changing resolution from 0.050 to 0.010:
The screen shown above appears once change of resolution is requested from the
SETUP
WEIGHING
screen by moving the cursor using and so the
Resolution
parameter is se-
lected and then pressing
.
In order to change the “Resolution” parameter
to 0.010 perform
the following:
•
Press and/or repeatedly until
0.010
is selected in the list.
•
Press
to accept the selection.
Entry of numbers
Some parameters (such as minimum and maximum weight of the weighing range) are en-
tered using a data entry screen. When change of this type of parameter is requested, a data