Aquameter
™
Series Instruction Manual
10401-00188 Rev M
16.4. Uploading Data From Your Aquameter
Ensure your Aquameter has batteries installed but is switched off. Connect the Aquameter
to your PC using the USB cable supplied. The Aquameter should switch itself on
automatically and display ‘USB CONNECTED’ on its screen.
Click the ‘
Upload Data From Aquameter’
button. AquaLink will search for the Aquameter
then upload all the available logged data from the Meter to your PC. A progress bar and file
counter will be displayed during this process.
Once upload is complete, the memory Tag, date and time for all the logged data that has
been uploaded will be displayed in the
Uploaded Data
column on the left of the screen.
To view any of the logged data records, simply click on the desired Tag, date and time label
as shown above. The data for the highlighted label will be displayed in the individual data
boxes, which are grouped by electrode function. Any data that is unavailable or out of
range will be displayed as dashes. To move up and down the Tag/date/time column, use
either your mouse or the cursor up/down keys.
Remember, the Aquameter stores all logged data in a raw Probe format, so can be made to
output logged data in several different forms, dependent upon the Meter’s current settings.
See
Important Information About Memory Mode
in section 8 for more information.
16.5. pH/ORP Electrode Data Logged Using an AP-100, AP-400, AP-500 or AP-600
If your data was logged using an AP-100, AP-400, AP-500 or AP-600, the data displayed in
the pH/ORP group of data boxes depends upon the setting in the Meter. If the Meter is set
to display pH or pHmV, the
pH
data box will display pH, the
pHmV
box will display pHmV
and the
ORP
box will display dashes. If an ORP electrode was fitted to the Probe and the
Meter is set to display ORP, the pH box and pHmV box will display dashes. The box
labelled
ORP
will display ORP in mV. See section 9:
more details.
© 2011 Aquaread Ltd.
www.aquaread.co.uk
Page 45 of 64