6.5.3 Recommended angle and default exposure time for each tooth type
Receptor
Patie
nt
Teeth
Angle of
inclination(The
patient sits in a
chair
vertically.)
Exposure
time(s)
Digital
Sensor
Adult
Bitewing
+5˚~ +8˚
0.4
Lower
Molar
-
5˚
0.2
Upper
Molar
+30˚
0.32
Canine
Maxilla: +45˚
Mandible: -
20˚
0.25
Incisor
Maxilla:
+45˚
Mandible: -25
˚
0.16
Child
Bitewing
+5˚~ +8˚
0.32
Lower
Molar
-
5˚
0.13
Upper
Molar
+30˚
0.2
Canine
Maxilla: +45˚
Mandible: -
20˚
0.1
Incisor
Maxilla: +45˚
Mandible: -25
˚
0.08
Phosphor
Plate
Adult
Bitewing
+5˚~ +8˚
0.5
Lower
Molar
-
5˚
0.25
Upper
Molar
+30˚
0.4