Eighteeth Fast-Pack pro User Manual Download Page 1

 

                  

 

 

 

Fast-Pack pro

 

USER MANUAL 

Summary of Contents for Fast-Pack pro

Page 1: ...Fast Pack pro USER MANUAL...

Page 2: ...P N IFU 6335007 Version 01 Issued Nov 04 2020 Size 85mm 119mm...

Page 3: ...Pack pro 10 4 1 Installation of heating needle 10 4 2 Installation of adapter 10 4 3 Connecting charge base 11 5 Use Interface 12 6 Setting 13 6 1 Memory Parameter Setting 13 6 2 Advanced Setting 14...

Page 4: ...1 Scope of Fast Pack pro 4 31 1 Scope of Fast Pack pro 1 1 Parts Identification 1 2 3 4 1 Charge Base 2 Handpiece 3 Heating Needle 3PCS 4 Adapter...

Page 5: ...essories Fast Pack pro Handpiece 1pcs Charge Base 1pcs Adapter 1pcs Heating Needle S 1pcs Size 40 0 025 Color Black Heating Needle M 1pcs Size 50 0 05 Color Yellow Heating Needle L 1pcs Size 60 0 06 C...

Page 6: ...he user patient NOTE Additional information explanation of operation and performance Serial number Catalogue number Manufacturer Date of manufacture LOT Lot of manufacture Class II equipment Type B ap...

Page 7: ...ser Manual 7 31 Community 55 C 20 C Temperature limitation 80 20 Humidity limitation 106kPa 70kPa Atmospheric pressure limitation Manufacturer s LOGO Consult instructions for use Washer disinfector fo...

Page 8: ...ce must be operated and stored in a safe environment 3 The device requires special precautions with regard to electromagnetic compatibility EMC and must be installed and operated in strict compliance...

Page 9: ...ll of the leaked liquid and contact the local agency 9 Do not use this device for any dental procedure other than root canal obturation 10 Thermal hazard risk exists for patients Cautions should be ex...

Page 10: ...ce then can be installed in other orientations WARNING After the operation is completed wait for the heating needle to cool down and remove it to prevent the heating needle from being damaged accident...

Page 11: ...the original adapter could be used Put the handpiece all the way into the charge base the charge state will show on the screen NOTE Put the handpiece into the charge base in the right direction otherw...

Page 12: ...ameter Setting During standby state holding down press then press for 2 seconds to enter memory parameter setting Parameter T1 to T5 can be set independently Heating Long press Turn Power Off Long pre...

Page 13: ...is memory can be changed Press till target temperature the temperature can be set to 90 120 140 150 160 180 200 220 250 and 300 Press to confirm Keep heat Time Sec 10 Press again the Keep heat Time of...

Page 14: ...tes Press to confirm Beep Volume Vol 1 Press again the Beep Volume can be changed Press to adjust the Beep Volume can be set to 0 1 and 2 Press to confirm RestoreSettings NO YES Press again the Restor...

Page 15: ...ended See chapter 4 3 Connecting charge base NOTE Only the original adapter could be used Do not use the device while charging The handpiece power connector can only be used to connect the original ad...

Page 16: ...the gutta percha 1 2 3 166 2 1 The Keep heat Time will display on the screen When the set time is reached the heating process will stop 2 Heating indication 3 Real time heating temperature 2 1 66 Rel...

Page 17: ...sing The device should no longer be reused in case of signs of material degradation In case of damage the device should be reprocessed before sending back to the manufacturer for repair 8 2 General re...

Page 18: ...uments in a humid surrounding Do not submerge the components or wipe them with any of the following functional water acidic electrolyzed water strong alkaline solution or ozone water medical agents gl...

Page 19: ...ermediate rinsing with warm water 40 C Emptying The automated cleaning processes have been validated by using 0 5 neodisher MediClean forte Dr Weigert Note Acc to EN ISO 17664 no manual reprocessing m...

Page 20: ...ts have been maintained according to the manufacturer s instruction Packaging Pack the instruments in an appropriate packaging material for sterilization WARNING Check the validity period of pouch giv...

Page 21: ...iod Reprocessi ng validation study information The above mentioned reprocessing process cleaning disinfection sterilization has been successfully validated Refer to cleaning disinfection validation re...

Page 22: ...ghtly moistened with Ethanol for Disinfection Ethanol 70 to 80 vol at least 2min repeat for 5 times NOTE Do not use anything except Ethanol for Disinfection Ethanol 70 to 80 vol Do not use too much et...

Page 23: ...not connected Check the connection 4 3 The plug of the adapter is not inserted into the outlet Check the connection There is no electricity in the outlet Check the connection No charge indicator flas...

Page 24: ...0 Charger power output 6V 3A Frequency 50 60Hz 10 Power rating 15VA Temperature 90 300 Electrical safety class Class II Applied part B Operation mode Continuous operation Operating conditions Use in e...

Page 25: ...uding domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Class A Vol...

Page 26: ...at 0 0 UT 250 300 cycle 0 UT 0 5 cycle at 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles sine phase at 0 0 UT 250 300 cycle Mains power quality should be that of a typical commer...

Page 27: ...separation distances 3 V 3V m Complies Portable and mobile RF communications equipment should be usedno closer to any part of the Fast Pack pro including cables than the recommended separation distanc...

Page 28: ...dulation 18Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN...

Page 29: ...ectromagnetic immunity of Fast Pack pro and result in improper operation Cable information Cable Name Cable Length m Shielded or not Remark Adapter Cable 2 No 2 Use of Fast Pack pro adjacent to or sta...

Page 30: ...ical components and printed circuit boards are disposed as electrical scrap The lithium batteries are disposed as special refuse Please deal with them according to the local environmental protection l...

Page 31: ...ad Xuejia County Xinbei District Changzhou City 213000 Jiangsu P R China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www eighteeth com Caretechion GmbH Tel 49 211 3003 6618 Add...