4.1
Installation of Gutta-
percha
Push the Gutta-percha into the
handpiece.
NOTE
This product is not included the Gutta-
percha. Please use the Gutta-percha
that is recommended by Sifary. Refer
to the Technical data of this manual for
the recommended Gutta-percha size.
4.2
Installation of Cartridge
Screw the cartridge along the thread of
the handpiece to install.
The needle on cartridge cannot be
turned or bended by hand.
Use bending tool to bend the needle.
Use bending tool with the hexagonal
hole matched to rotate the needle.
Unscrew the cartridge from the
handpiece to disassemble.
WARNING
➢
To remove the cartridge. As shown in
the picture, touch the flow heater
with finger rapidly, make sure it is not
hot before operation. If the
temperature is too high, waiting 3~5
minutes till it is cooling.
➢
Even if the cartridge cools down
already, we strongly recommend not
to touch the needle on cartridge.
There is a risk of heat injury or
damage the cartridge. Hold the black
shell to remove.