![EG Xcellerex XDUO 2500 Operating Instructions Manual Download Page 232](http://html1.mh-extra.com/html/eg/xcellerex-xduo-2500/xcellerex-xduo-2500_operating-instructions-manual_2367561232.webp)
Action
Step
Verify that the clamp on the drain line is fully engaged.
3
Remove the two side access doors, if not already removed.
4
Follow the instructions below to remove the doors:
a
Hold the door handle tightly.
b
Turn the door counterclockwise.
c
Remove the door.
Connect the liquid source to the addition line of the bag assembly.
5
If your mixer has a weight monitoring option, tare the mixer weight. See
for instructions.
6
Add liquids into the disposable
bag
CAUTION
Check floor safety limits. Before filling the TCU and the mixer with
liquid, verify that the total system weight is below the established
floor loading safety limits of the facility.
232
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
7 Operation
7.8 Add liquid contents