79
8.4.3
File Editing Commands
The following commands are used to create or modify Test Setup Files.
Commands should be separated from parameters by a space.
If multiple parameters are entered, they should be separated by commas.
COMMAND
DESCRIPTION
VALUE
FL <memory number>
File Load
memory number = 001-200
FN nn,xxxx
File Name
nn=01-2000, xxxx=name
FD
File Delete
Current file in use
FD nn
File Delete nn
nn=1-200
FS
File Save
Current file in use
FSA nn,xxxx
File Save As
nn=1-200, xx=Name
FT?
File Total
ST?
Step Total
SS <step number>
Step Select step
number
step number = 1-200
SS?
Step Select ?
SAA
Step Add ACW test
SAD
Step Add DCW test
SAD-
Step Add DCW- test
SAI
Step Add IR test
SAG
Step Add GND test
SAC
Step Add CONT test
ADD2 <n
,p1,p2,p3…>
Add all parameters of
a test step
n=test type, p=parameters
ADD3 <n
,p1,p2,p3…>
Add all parameters of
a test step
n=test type, p=parameters
SD
Step Delete current
SD nn
Step Delete nn
nn=01-200
SP <prompt message>
Step Prompt Create
prompt message = Valid ASCII (1)
maximum 32 characters
SP
Step Prompt Delete
LP?
List Prompt
LP nn
List Prompt current
(1) "Valid ASCII" is the character set that is available from the front panel LCD user interface.
Consisting of upper case alphabet (A-Z), numbers (0-9) and decimal point (.), asterisk (*), dash (-),
under bar (_), tilde (~) and space (SP).
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory RAM.
FN < file name>
Creates a new file name for the active memory loaded into RAM.