39
the exact alpha-numeric code that is on the product’s barcode label. For example, if Product A has
barcode “123456789”, then the test file that the user would like to run when testing Product A
must be named “123456789”. Upon scanning the barcode, the SE will immediately load the test
associated with that barcode. The test file name is limited to 10 characters. However, if the user
names a test file with the maximum 10 characters, this function will still initiate a test when a
product’s barcode begins with those first 10 characters, even if the barcode has more than 10
characters.
When using certain features of barcoding, the instrument’s response to the TD? and RD x?
commands will differ slightly to an instrument that is not utilizing this function. For all types of
tests (ACW, DCW, IR, CONT) two fields are added to the end of the standard response when the
Barcode I/P setting is set to SERIAL#, PRODUCT# or SER/PROD.
The first field contains the Serial Number information and the second field includes the Product
Number information. Both fields are included regardless of which of these three modes are
selected. The Data Storage Card will simply substitute a “0” for the field if it is not applicable to the
setting. For example, if a user had their Barcode I/P setting set to SERIAL#, and scanned a Serial
Number with the value “123456789”, the TD? response for an ACW test could be:
01,ACW,Pass,1.24,1.000,0.900,1.0,123456789,0
Note that there is a “0” in the Product Number field because the Barcode I/P setting is SERIAL#.
When the Barcode I/P setting is RUN FILE or OFF, these fields are not included in the TD? and RD
x? responses.
This is where the selection can be saved by pressing the enter key (
↵
)
and move to the previous
screen.
The Autostart function works in tandem with the Barcode RUN FILE function to automatically start
a test sequence once the product and serial number have been scanned.
Using the Autostart = ON and RUN FILE feature will enable the instrument’s
output once the barcode is scanned. Do not touch the DUT at any time
when using this feature in order to avoid potential shock or serious injury.
The Autostart function can be set to ON or OFF. To set the Autostart function touch the ON/OFF
key. To save the setting touch the enter key (
↵
)
and move to the previous screen.
4.3.3.15 Result Data
From the Setup System menu touch the Results Setup icon and the following screen will appear: