Installation
A732-01-880 Issue A
3.3
Unpack and inspect
Remove all packing materials, remove the pump from its packing box, remove the protective covers from the inlet
and outlet ports and inspect the pump. If the pump is damaged, notify your supplier and the carrier in writing; state
the item number of the pump together with your order number and your supplier's invoice number. Retain all the
packing materials for inspection. Do not use the pump if it is damaged. If the pump is not to be used immediately,
replace the protective covers. Store the pump in suitable conditions, as described in
. Refer to
for disposal of materials.
3.4
Locate the pump
Provide a firm, level platform for the pump. Locate the pump so that the Run/Standby switch is accessible.
If your pump will be located inside an enclosure, make sure that there is adequate ventilation at both ends of the
pump, so that the ambient temperature around the pump does not exceed 40 ºC. There must be a minimum space
of 25 mm between the pump and the enclosure walls.
3.4.1
Mechanical fixing
Note:
If you wish to secure the pump you may do so using the four holes located on each corner of the pump base,
we recommend you use M8 bolts.
3.5
Electrical installation
3.5.1
Check and configure the pump
CAUTION
Before changing the voltage, unplug the power supply. Ensure that the pump is correctly configured for your
electrical supply otherwise you may cause damage.
Ensure that the voltage shown on the voltage indicator (refer to
, item 7) on the motor cover corresponds
with your electrical supply voltage. If it does not, you must change the configuration of the pump motor to match
your supply voltage; use the procedure below.
1. Undo the two screws and lift off the voltage indicator moulding.
2. Rotate the voltage indicator moulding so that the correct voltage is uppermost.
3. Replace the two screws.
WARNING
Use suitable lifting equipment to move the XDS100B booster. The mass of the pump is 48 kg.
z
Mechanical lifting equipment should be attached to the lifting eye; loose slings should not
be used.
z
Take care when moving the pump into position. Its weight makes it difficult to slide.
Movement should be attempted using two people.
WARNING
If the pump is to be used on the floor of a work area, position the power lead and the exhaust and
inlet hoses with care. Ensure that personnel in the area are aware of any obstructions around the
pump.