Page 7
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
A732-01-880 Issue A
2.3.2
Vibration data
2.4
Electrical data
Note:
Time-lag fuses should be used, as the initial inrush current can exceed rated values briefly when pump is
started.
Table 5 - Vibration data
Parameter
XDS100B
Sound pressure, measured at ultimate vacuum 1 metre from the end of the pump
to ISO 3744 and ISO 4871
52.1 +/- 2.5 dB(A)
Vibration: measured at the inlet port (ISO 3744)
Class 1C…< 4.5 mms
–1
(rms)
Table 6 - Electrical ratings for continuous operation
Pump
Supply (V)
Phase
Frequency (Hz)
Current (A)
XDS100B
200/230
Single
50/60
6.6
100/120
Single
50/60
10.6
Table 7 - Recommended fuses
Area
Voltage
Rating
UK
230 V
13 A
Europe
230 V
16 A
US
110 V
20 A
Japan
100 V
20 A