Figure 7 Valve connector showing pin numbers
Pin number
Signal
Polarity
Colour
1
Valve control output Positive
Brown
2
Not used
n/a
n/a
3
Return
Negative
Blue
4
Not used
n/a
n/a
3.10 Indicator LEDs
Table 18
Indicator LEDs
LED
Description
Service indicator LED
If a service is due, this LED flashes in a sequence to
indicate which service operation is required. Refer to
Service indicator codes on page 70
.
Alarm indicator LED
This red LED flashes in a sequence to indicate an error
code if a FAIL condition is preventing pump operation.
The error codes can be used for fault finding as described
in
Alarm indicator codes on page 76
.
3.11 Serial protocol
The serial interface link is set to 9600 Baud, 8 bits, 1 stop, no parity with no
handshaking. The commands are made up from printable ASCII characters. The
maximum message size that can be sent is 80 characters, including start and end
characters.
Note:
All alphabetical characters must be sent in upper case format. The response may contain
lower case characters.