UPDATE TO URGENT FIELD SAFETY NOTICE 

FCA #128  

Product:

 EV1000 clinical platform

 

Model Numbers:

 EV1000A, EV1000NI 

Serial Numbers: 

All serial number for models listed 

 

ACTION REQUIRED

 

<MM DD, YYYY> 

 

<Customer #> 

<Contact name or Dept.> 

<Firm Name> 

<Attention: RISK MANAGEMENT> 

<Address> 

<City/state/zip> 

 

Dear Valued Customer: 

 

This letter is in follow-up to our previous communications regarding EV1000 monitors and the risk 

associated with the power supply as communicated earlier. 

 

Edwards is working to address all the root causes, and to implement corrective and preventative 

actions as quickly as possible.  There are two types of monitors, each with different power supply 

configurations and different actions, as described below. 

 

Edwards would like to highlight the importance of: 

 

EV1000A 

The  EV1000A utilizes the FloTrac sensor  and is powered by an external power supply.  As 

previously noted, it is important that the “brick” of the power supply always be oriented correctly  

to avoid the risk of liquid ingress as shown below and according to the equipment operators’ 

manual.   

 

Correct Orientation 

 

Flag Label 

 

 

 

Summary of Contents for EV1000A

Page 1: ...s communicated earlier Edwards is working to address all the root causes and to implement corrective and preventative actions as quickly as possible There are two types of monitors each with different...

Page 2: ...al contaminants These visits are in progress and may have already occurred at your facility In the event your staff mistakenly fail to follow the instruction on the flag label we are providing a newly...

Page 3: ...stalled to avoid contamination with exception of activities per your policy for inspection and cleaning and in accordance with cleaning instructions in the Operator s Manual While the power cord entry...

Page 4: ...ack of the pump unit This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred If...

Page 5: ...rial number 1 EVDXXXXX and enter an X or Yes on Flag Label and Power Cover Adapter or indicate N A if the equipment already has any of the components or if you have no product o For EV1000NI or EV1000...

Page 6: ...y became aware of an issue with Edwards EV1000 clinical platform and are notifying our customers Details on affected devices The EV1000 clinical platform is used to optimize fluid management through t...

Page 7: ...Based on review of previous instances we strongly advise you to be cautious that IV bags and other fluids need to be positioned away from AC power connectors in order to avoid liquid ingress For EV100...

Page 8: ...3 of 6 Reference Warnings and Caution Section from Operators Manual Figure from EV1000A Operators Manual on Power Adapter Orientation Figures 15 8 and 15 9...

Page 9: ...ead to an electrical short circuit leading to a possible fire hazard due to liquid ingress Unlike the EV1000A the EV1000NI AC cord inserts horizontally into the unit Therefore orientation is not a fac...

Page 10: ...n where the potentially affected devices have been transferred If you have any questions please contact Edwards Customer Service or Tech Support at phone Sincerely Vice President of Quality Critical C...

Page 11: ...turn any product Verify your inventory and correct setup of the system Complete all sections of the table below indicate 0 if you have no product Fax e mail the completed form to Edwards Customer Serv...

Page 12: ...s to inform you of additional potential causes our plans to address the issues and steps you can take to help minimize risk Reports of Burnt Cables We have reports of cables that appear burnt without...

Page 13: ...ntamination by conductive materials prior to use completed with this notification Updating the Operators Manual to include the instructions as noted above and any others identified as our investigatio...

Page 14: ...structions below to complete the acknowledgement process Complete this acknowledgement form to confirm that you received this addendum DO NOT return any product Email the completed form to Edwards Tec...