Figure 8 Connections and direction of motor rotation
1. Pump body temperature control connector
2. Purge gas inlet
3. Cooling-water outlet
4. Cooling-water intlet
5. Exhaust
1. Pump body temperature control connector
2. Purge gas inlet
3. Cooling-water outlet
4. Cooling-water intlet
5. Exhaust
Type
Cooling-water demand at feed temperature(assuming a constant
discharge temperature of 50 °C)
30°C...35°C
25°C...30°C
5°C...25°C
l/min
l/min
l/min
EDS 200/300
8.0
6.0
4.0
Depending on your application and use of thermostatic valve temperatures and flows may differ
Figure 9 Cooling water level
CS/
4062/
A
max.
min.
7.4.1 Water quality
In order to ensure long trouble-free operation the cooling-water must not contain any oils,
greases and suspended solids. Moreover, we recommend compliance with the following
limit values:
Appearance
Clear, free of oils and greases
Suspended matter
< 250 mg/l
Particle size
< 150 μm
Electrical conductivity
< 700 μS/cm
pH value
7.0 to 9.0
Total hardness (total alkaline earths)
< 8 °dH
Aggressive carbon dioxide
None, not detectable
Chloride
< 100 mg/l
Sulfate
< 150 mg/l
Nitrate
≤ 50 mg/l
Iron
< 0.2 mg/l
Manganese
< 0.1 mg/l
Ammonium
< 1.0 mg/l
Page 28
A41802880_C - Installation