Operation
D397-53-880 Issue C
4.7.73
Module 1011 - CXS On Process Warning
CAUTION
Use of this option increases likelihood of pump stop during Process.
To avoid premature controller failure this parameter is should not be changed more than 100 times/day.
Normally a pump will not go On Process if a Warning exists. This module can be used to change a controller's
configuration to allow pump to go On Process even if the pump is in a Warning state.
Output Byte:
Values: 1=True (allows pump to go on Process despite Warning) 2=False (sets normal state which
inhibits pump going On process if a warning exists); Other values - ignored.
4.7.74
Module 1012 - CXS On Process Cold
CAUTION
To avoid premature controller failure this parameter is should not be changed more than 100 times/day.
Normally a pump will not go On Process unless warm. This module can be used to change a controller's configuration
to allow pump to go On Process even if the pump is cold.
Output Bytes: Values: 1=True (allows pump to go on Process even if cold) 2=False (sets normal state which inhibits
pump going On process if pump is cold); Other values - ignored.