Start/Stop a Measurement and View Arterial Waveform

1.   Zero the HRS sensor (see Zeroing the HRS).

2.   Tap the 

Pump Unit Start/Stop

 icon on the navigation bar 

(1)

.  

An arterial waveform will appear at the top of the screen 

(2)

.

3.   Press the 

Home

 button to confirm.

4.   You may also view the arterial waveform on the trend screen by  

tapping the 

Screen Selection

 icon on the navigation bar, selecting  

the 

Trend Screen

 icon, and tapping the 

Arterial Pressure  

Waveform

 icon

 

(3)

.

5.   Use the same icon to remove the waveform

 

(3)

.

6.   To stop a measurement, tap the 

Pump Unit Start/Stop

 icon on  

the navigation bar

 

(1)

.

3

2

1

1

2

3

Applying the Finger Cuff

1.   Determine the correct finger cuff size  

by referring to the EV1000 software  

under Settings>Help>ClearSight> 

Device Setup>Step 3.

2.   Place the index, middle, or ring  

finger in cuff #1 between both  

optical components and center  

between the two knuckles 

(1)

.

3.   Gently lead the cuff cable and air  

hose between two fingers toward  

the back of the hand, as shown  

in the picture on the right.

4.   Wrap the finger cuff snugly around the contour of the finger  

ensuring that the end of the finger cuff lies inside the green zone 

(2)

. The image of the finger should lie on the top side of the finger.

5.    Connect the finger cuff to the pressure controller 

(3)

.

6.   If using a second finger cuff,  

repeat steps 1-4 on adjacent  

finger with cuff #2.

Applying the Heart  
Reference Sensor (HRS)

1.   Secure the pressure controller  

to the Velcro strap with the cuff  

connections facing the fingers.  

Wrap the Velcro strap around the  

wrist and secure it 

(4)

.

2.   Connect the HRS to its respective  

port on the pressure controller 

(5)

3.   Secure the finger side of the HRS to one of the finger cuffs 

(6)

.

4.   Secure the heart side of the HRS to the patient at heart level 

(7)

.

NOTE: 

Use surgical tape to properly secure both  

ends of  the HRS if necessary.

6

4

7

5