TECHNICAL DATA
2.3
Vent gas specification and vent control data
Although the pump may be vented to atmosphere, high relative humidity of the air may greatly increase the
subsequent pump-down time. To reduce pump-down times, vent the pump with dry, clean gases.
Vent gas
Dry air, nitrogen, argon or other inert gases
Maximum dew point at atmospheric pressure
-22 °C
Maximum size of particulate
1 µm
Maximum concentration of oil
0.1 parts per million
Time for rotational speed to reach 50%
>15 sec
Table 1 - Technical data
Parameter
EXT70/70/200H
Notes
Mass (system)
>23 kg
Eyebolt positions are supplied
for lifting and handling
Mass (Cartridge pump)
9 kg
Outlet flange
DN25NW
Main inlet pumping speed (TOF)
N2
*
170 ls
-1
187 ls
-1
164 ls
-1
169 ls
-1
Front-side inlet pumping speed (Transfer chamber)
N2
110 ls
-1
153 ls
-1
100 ls
-1
H
2
133 ls
-1
Back-side Inlet pumping speed (Tunnel optics)
N2
177 ls
-1
167 ls
-1
167 ls
-1
H
2
103 ls
-1
Maximum continuous inlet pressure
Air cooling at 35°C
†
Main inlet
1.6 x 10
-2
mbar (1.6 Pa)
Air cooling is beneficial to the
EXDC160 controller
Transfer
3.36 x 10
-2
mbar (3.36 Pa)
Tunnel optics
3.62 x 10
-2
mbar (3.621 Pa)
Nominal rotational speed
60000 r min
-1
Starting time to 90% speed
EXDC160
115 sec
Noise level (at 1 metre)
< 50 dB(A)
Recommended controller
EXDC160 24 V
EXDC160 maximum VA input
250 VA
Quiescent power
25 W