background image

 

© 2014 UTC Fire & Security Americas Corporation, Inc. 

1 / 6 

P/N 3100672-EN • REV 07 • ISS 23JAN14 

 

AB4G Audible Detector Base Installation Sheet

 

Description  

This installation sheet applies to AB4G Audible Detector Base 
units with model numbers GSA-AB4G and SIGA-AB4G. 

The AB4G Audible Detector Base adds an audible output 
function to any Signature Series detector. The output of this 
detector base is field-configurable for output tone (steady or 
temporal) and output volume (low or high dBA). 

Depending on the system supporting the Signature loop, the 
base can: 

• 

Follow the state of the device it supports 

• 

Be configured (in the SDU) for other operating modes 

• 

Be controlled by program rules 

The base uses the same address and programming label as 
the detector it supports. 

Installation 

Install this device in accordance with applicable national and 
local codes, ordinances, and regulations. 

Cautions 

• 

To avoid accidental damage to the panel, disconnect all 
power before wiring the unit. 

• 

Do not loop the signaling circuit field wires around the 
terminals. 

Note: 

Always connect the base to a continuous voltage, 

whether the output tone on the audible detector base is set to 
steady or temporal. 

 

Sleeping rooms: 

In sleeping areas, use the high dBA output 

and temporal tone settings.  

AB4G-SB: 

When using the AB4G-SB box, install a reinforcing 

plate at every knockout. (Reinforcing plates are included with 
the box.) Remove the knockout first and then slide the 
reinforcing plate into the plastic housing. After the plate is in 
place, install the conduit connector and nut. See Figure 1.  

Typically, the base is configured to produce a high dBA 
temporal tone and is connected to a notification appliance 
circuit that outputs a continuous 24 VDC signal. 

Refer to Figure 2 when following the installation steps given 
below. 

To install the audible detector base: 

1.  The unit default is for high dBA output. To set the output to 

low dBA, cut the circuit board trace as marked on the back 
of the PC board. See Figure 3. 

2.  The unit default is for temporal pattern output. To set the 

output to steady tone, cut the circuit board trace as 
marked on the back of the PC board. See Figure 3. 

3.  Select and install a compatible electrical box, and then 

bring the field wiring into the box. 

4.  Connect the field wiring to the terminals on the back of the 

base plate. For the unit to function properly, observe 
polarity. See Figure 3. 

5.  Attach the base plate to the electrical box. 

6.  Align the trim ring so that the four tabs on the ring mate 

with the four slots in the base plate, and then press the 
trim ring onto the base plate until the tabs lock. 

7.  Attach the desired Signature detector to the base. Align 

the arrows on the detector and trim ring, press the 
detector into the base, and then rotate the detector until it 
locks into place. 

8.  Apply power and activate the unit to verify that it is 

operating properly. 

Reviews: