Storage and disposal
B064-04-880 Issue F
6
Storage and disposal
6.1
Storage
Use the procedure below to store the pump:
1. Shut down the pump as described in
.
2. Isolate the pump from the electrical supply, disconnect it from the vacuum system and drain the pump fluid as
described in
.
3. Dismantle, clean and reassemble the pump as described in
. Do not refill the pump with fluid.
4. Place protective covers over the inlet and outlet-flanges and the cooling-water connection points.
5. Where possible, store the pump in cool dry conditions until it is required for use. When required, prepare and
install the pump as described in
.
6.2
Disposal
Dispose of the 30B5M pump and any components removed from it safely in accordance with all local and national
safety and environmental requirements.
Take particular care with the following:
Components which have come into contact with pump fluid.
Fluoroelastomers which may have been subjected to temperatures above 260 ºC (see
).
Components which have been contaminated with dangerous process substances.