TEMPERATURE CALIBRATION
1. Use the DS299S to remove the temperature
control knob. Remove RS321 sleeve. Turn
temp. control (A) so arrow points to cord.
2. Use the SRW01 to pull out heater assembly,
PCB & power cord from handle. Leave
enough power cord slack to install SRF30.
3. Place heater assembly, PCB & power cord
inside SRF30. Install RS321 and connect to
air source.
4. Follow set-up shown using of the MS412
Temperature Calibration System. Includes:
•TI680
•SDS100
•FX635
5. Insert the hot air tip inside the TPL09-1,
place the center of the thermocouple wire of
the SDS100 inside the slot of TPL09-1 loca-
tor. Use only CLEAN thermocouple wires.
6. Turn on air, adjust to 5 scfh airflow.
7. Turn temp. control (A) fully clockwise. Adjust
calibration pot (B) until reading stabilizes at
850°F.
8. Assemble unit in original handle.
SRW01
DS299S
SD418
TPL09-1
Items Needed
• SRW01
• DS299S
• SRF30
• MS412
• TPL09-1
• SD418
Calibration Pot (B)
Temp. Control (A)
SRF30
original handle
Hot Air Tip
TPL09-1
Ceramic
Locator
FX635
TI680
SDS100