LYNCS-WI Operation
1.
Select the amount of toolposts to match the number of wedges / probes being deployed.
2.
Remove one tractor assembly.
3.
Slide on any cable management guides required.
4.
Arrange tractor units and affix toolposts depending on required application.
5.
If mounting pins required, select the appropriate and screw on to toolpost arms.
6.
Load probe mounted wedges to scanner.
7.
Insert encoder cable and attach strain relief to strain relief point.
8.
Attach cable management sleeve to strain relief points.
9.
Set toolpost in desired position along rail and lock into place. If utilising outer toolposts do not set
for pipe diameter until on the test item.
10.
Tether scanner from both handles.
11.
Setup data acquisition unit
–
for M2M instruments select LYNCS-WI in scanner list NOTE: Typical
encoder resolution: 25.85 pts/mm.
12.
Perform encoder calibration and complete remaining setup configuration.
13.
Place onto test item to perform scan.
If using M2M instrument, when setup is complete, press the
Play
button immediately followed by the
Pause
button. Scanner is now ready to make use of the on-board control Pause /
Resume
function.