eDAQ Potentiostats
recognised the Potentiostat. Exit the software, check all your
connections and try again).
By default, the control window opens with the Potentiostat in Standby
mode, that is with the reference and working electrodes isolated so that
no current will flow through your electrodes. To connect to the
Potentiostat lead wires you must select Real mode. When you click
Cancel or OK the Potentiostat will revert to Standby mode until
recording is started.
Now select Dummy mode operation. You will need to adjust the gain
range to 20 μA to accommodate your signal amplitude. You can now
adjust the applied potential with the slider bar, or by entering the exact
potential with the A-button. The resulting current signal should obey
Ohm’s law:
I = E/R
so that an applied potential of 1 V should produce a current of 10 μA,
while other potential settings should produce corresponding currents.
Select Potentiostat in the
Channel pop-up menu
Signal display area
Drag ticks and
labels to adjust axis
Figure 2–7
Potentiostat controls with
Chart software.