VE-H100B Veterinary Pulse Oximeter User Manual
- 45 -
Alarm
, there will be
↑
or
↓
icon displayed on screen, and the SpO
2
or PR character will flash.
If
Alarm system
is set to
OFF
, all audio alarms and visual alarms
are turned off.
WARNING
When the Alarm system is off, the oximeter will not give an
alarm prompt. In order to avoid e
ndangering the patient’s life,
the user should use this function cautiously.
3.3.10 SpO
2
Alarm Setup
The user can choose
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
in
menu to adjust SpO
2
alarm limit. Press the
Up Arrow
button or
Down Arrow
button to increase or decrease alarm limit.
By default,
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
are
100
and
85
respectively.
Set the SpO
2
alarm limits as follows:
Choose
SpO
2
Hi Alarm
in the menu, press the
Function
button to enter setup. The
SpO
2
Hi Alarm
box will change from real line box to broken line box.
The adjustable range for upper limit of SpO
2
is from
“1 + the lower limit of SpO
2
” to 100. If the value of
SpO
2
Hi Alarm is set to less than 85, it will restore to