SE-601 Series Electrocardiograph User Manual System Setup
- 42 -
Chapter 10 System Setup
Select
Setup
on the main screen1 to display the
System Setup
screen.
10.1 Work Mode Setup
Items
Default
Default
(Only in the U.S.)
Mode Options
Auto, Manual, Rhythm
Auto, Manual, Rhythm, R-R
Display Style
6×2
3×4+1R
Sampling Mode
Real-time Sample
Real-time Sample
Rhythm Style
Three Leads
Three Leads
Duration
(Periodic Sample)
60 min
60 min
Interval
(Periodic Sample)
1 min
1 min
Preview
Off
Off
Auto Arrhythmia Detection
Off
Off
10.2 Filter Setup
Items
Default
Default
(Only in the U.S.)
AC filter
On
On
EMG filter
Off
Off
DFT filter
0.67Hz
0.67Hz
Lowpass filter
100Hz
100Hz
SE-601A: CMRR ≥ 121dB (when AC filter is ON)
SE-601B/C: CMRR ≥ 140dB (when AC filter is ON)