Principle Introduction
- 21 -
Charge interface board ports to indicative light board
Figure 7- 6
7.1.4 Power module
Power module outputs
+
12V and
+
5V and is responsible for charging management.
When the system is supplied by 90-264VAC power, power module transfers AC mains into 12V
DC and 5V DC power. When the system is cut off from the 90-264VAC powe, the power
modules transfer the battery power inside into 12V DC and 5V DC power.
Pin No. Symbol
1,2,3,4 VCC
5,6,7,8 GND
9,10
TRANS1_TX
11,12
TRANS1_RX
13,14
TRANS2_TX
15,16
TRANS2_RX
17,18
TRANS3_TX
19,20
TRANS3_RX