FTS-3 Fetal Telemetry System Service Manual
Maintenance
- 15 -
Chapter 6 Maintenance
6.1 Maintenance Inspection
(1) Visual Inspection
Please perform the following check before you use FTS-3:
Check the base station and transducers to see if there is any visible damage that may
affect patient’s safety.
Check all the outer cables, power socket and power cables.
Check if the system functions properly.
If any damage is detected, replace the damage part(s) or contact EDAN immediately.
(2) Routine Inspection
The overall check of FTS-3, including the safety check and functional check, should be
performed by qualified personnel every 6 to 12 months, and each time after service.
The equipment should undergo periodic safety test to insure proper patient isolation from live
parts. This should include leakage current measurement and insulation testing. The recommended
testing interval is once a year or as specified in the institution’s test and inspection protocol.
Ground Leakage Current (Limit)
N.C. S.F.C.
500µA 1000µA
Enclosure Leakage Current (Limit)
N.C. S.F.C.
100µA 500µA
Patient Leakage Current (Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 500µA
Patient Auxiliary Current (Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 500µA
Patient Leakage Current (Limit)
DECG, MHR(from MECG)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Patient Auxiliary Current (Limit)
DECG, MHR(from MECG)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA