![EDAN FTS-3 Service Manual Download Page 20](http://html1.mh-extra.com/html/edan/fts-3/fts-3_service-manual_2357166020.webp)
FTS-3 Fetal Telemetry System Service Manual
System Configuration
- 14 -
Chapter 5 System Configuration
The end users can not change the system configurations. As a service engineer, you need to
change these configurations for them after FTS-3 is installed and checked properly.
NOTE:
Restart the base station after changing the settings.
5.1 Adjusting the Working Channel
If the fetal heart sound is with interference or it cannot be played smoothly, the working channel
is probably interfered. Put all the transducers back in the docking slots and press the adjustment
button
in back of the base station. The channel range is 1-14.
Restart the system when it enters the charging interface.
NOTE
:
The working channel number used by a system cannot be duplicated with that used by a
device of the same type.
理邦保密文件
EDAN CONFIDENTIAL