F15 Series Fetal & Maternal Monitor Service Manual
Troubleshooting
- 37 -
7.13 TEMP Monitoring Failures
Phenomenon
Possible Cause
Solution
No TEMP numeric value.
TEMP transducer defective.
Replace the TEMP transducer.
Bad connection of TEMP
transducer.
Check the connection.
iM X12 parameter board
defective.
Replace the iM X12 parameter board.
Main
control
board
defective.
Replace the main control board.