Cadence Fetal Monitor User Manual
- 46 -
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the
Cadence Fetal Monitor
is used exceeds the applicable RF
compliance level above, the
Cadence Fetal Monitor
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the
Cadence Fetal Monitor
.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.