TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 68 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
6.1
5.3.16 Converter efficiency
In case of measuring systems with a converter, the converter efficiency shall be at
least 98 %.
6.2
Equipment
Not applicable here.
6.3
Testing
Due to the measurement principle the tested measuring system does not use a converter.
6.4
Evaluation
Not applicable here.
6.5
Assessment
This test item does not apply as the measuring system does not use a converter.
Does this comply with the performance criterion? Not applicable
6.6
Detailed presentation of test results
Not applicable in this instance.